portable digital dental x-ray machine
Getting high quality, as well as clear X-ray images of the patients’ teeth, is vital for the dentist next step of treatment, the portable digital dental x-ray machine enables to bring the X-ray available when patient sit or lie in a dental chair unit that can be especially helpful in these situations. The advantage of a portable digital dental x-ray machine is that one portable dental device can be used in multiple dental practices which are considerably cost-saving compared to conventional wall-mounted dental x-ray machines.
The diagnostic portable digital dental x-ray machine is a device for photographing the teeth for accurate diagnosis. It aims to realize high-quality images and reduce unnecessary radiation exposure, and is a light portable dental device that can be operated by one hand.
Environmental and working conditions
- a. Condition of transportation and store
Environmental temperature: 10~40°C
Relative humidity:5%-90% RH
- b. Conditions of operation
Environmental temperature: 10~30°C
Relative humidity:10%-85% RH
Classification/Specification Compliance
- MDD(93/42/EEC):AnnexIX[Class II b]
- IPX specification[IPO: do not operate under wet conditions]
- Type of protection against electric shock[Internal powered device]
- Degree of protection against electric shock[Type B applied part]
- For use in environments where no flammable anesthetics and/or flammable
- Cleaning agents are present;non-alcohol based disinfectant only-wipes or cloth
- Dampened with liquid/spray.
Technical specifications
- Rechargeable Lithium Polymer battery:24V
- Maximum battery charge:25.2V
- Battery current at 2mA,60KV output:10A
- Power supply:JBL7451250750100FJ,Bridge Power Corp.
- Input:AC100-240V,50-60HZ 1.0A
- Output:DC25.2V,0.75A
- Cable length of DC output:1550mm
- Cable length of AC input:1550mm
- Cable length of remote controller:3000mm
X-Ray control and Generator
Exposure time range:0.01-1.6sec
Maximum duty cycle:1:60
Maximum inherent filtration:1.6mm Al
Maximum output power:120W nominal at 60KV,2mA
Generator rating:2mA at 60KV±5%
Leakage technique factors:60KV,2mA,1.6sec
Collimator cone
- Minimum source to skin distance:10cm,20cm(long cone)
- X-ray field size and configuration:6cm diameter circle
- Shielding:led-line
- GW:5.5Kg/set;
- Packing size:33*28*28cm